Product/Service

Custom Hi-Voltage Filtered Medical Connectors

Source: Quantic Corry

By applying high-voltage resistive ink banding using a patented banding process, these medical connectors are able to withstand the 5,000 VDC hi-pot breakdown required for defibrillation monitoring and medical diagnostic equipment. 

Materials:

  • Filter Elements: Ceramic
  • Housing: Tin Plated Aluminum
  • Leads: Gold plated phosphor Bronze, .025 + .001 diameter
  • Plastic Components: Zytel HTN FR52G30, Black
  • Encapsulant: Xiameter RTV 3110, White

Electricals:

  • Capacitance: 50 pf + 20%, PI circuit
  • DF: 2.5% Max
  • Insulation Resistance: 10 GOHMS Min @100 VDC
  • Dielectric Withstanding Voltage: 5,000 VDC, 2 SEC

Additional information can be found on the datasheet.