Monoblock by Dynarad is an innovative technology that integrates an X-ray tube and a high voltage generator into one physical assembly or subsystem. Monoblock is suitable for all x-ray applications, including medical x-ray (radiographic and flouroscopic imaging), analytical x-ray (XRD, XRF techniques), NDT or industrial x-ray analysis, or EDS (explosive detection systems). The Dynarad Division of Del Global Power Conversion Group developed Monoblock technology in order to reduce the cost and size of traditional tube plus generator configurations, while eliminating the high voltage cabling normally connecting these two components. As a result, Monoblock is more compact, more cost-effective and more reliable than traditional design approaches. Quality manufactured in our own ISO 9001 and FDA approved facility, Monoblock by Dynarad is the answer to today's rigorous X-ray applications that demand the best possible cost-size-performance ratio.
Window Type: Glass, Beryllium
Target Angle: 10º to 35º
Focal Spot: Varies from 10 micron to 2mm.
KVP Range: 20kV - 160kV
Input Power: 24VDC, 115VAC / 230VAC
Emission Current Range: 20µA - 200mA (pulsed)
Duty Cycle: CW or Pulsed
Line Regulation (KVP): ± 0.1% for a ± 5% line change (typical)
Line Regulation:
(Emission Current) ± 0.5% for a ± 5% line change (typical)
Load Regulation (KVP): ± 0.1% for load changes (typical)
Load Regulation:
(Emission Current) ± 0.5% for kV changes (typical)
Ripple: 1% RMS
Temperature Stability: 0.01% / ºC over the range of 10º to 35º (typical)
Set-ability: ± 0.5%
Rise Time: < 10mS for certain applications
Safety, Protective, and Interlock Features:KVP over-voltage fault
Emission current fault
KVP reference voltage limiting, adjustable
Emission current reference voltage limiting, adjustable
Fuses, circuit breakers, polyswitches
Optional heat exchanger
Optional collimator
MONOBLOCK SPECIFICATIONS
X-Ray Tube:
Single or double ended. Tubes manufactured by Oxford, Lohmann, Superior, Comet, Toshiba or customer's choice.Window Type: Glass, Beryllium
Target Angle: 10º to 35º
Focal Spot: Varies from 10 micron to 2mm.
KVP Range: 20kV - 160kV
Input Power: 24VDC, 115VAC / 230VAC
Emission Current Range: 20µA - 200mA (pulsed)
Duty Cycle: CW or Pulsed
Line Regulation (KVP): ± 0.1% for a ± 5% line change (typical)
Line Regulation:
(Emission Current) ± 0.5% for a ± 5% line change (typical)
Load Regulation (KVP): ± 0.1% for load changes (typical)
Load Regulation:
(Emission Current) ± 0.5% for kV changes (typical)
Ripple: 1% RMS
Temperature Stability: 0.01% / ºC over the range of 10º to 35º (typical)
Set-ability: ± 0.5%
Rise Time: < 10mS for certain applications
Safety, Protective, and Interlock Features:
Regulatory and Safety Standards:
Wherever applicable, UL, CE, and CUL components, standards, etc. will be utilitized.