Production Safety Analyzer
0.1µA to 9.999 mA.
Current trip levels for the AC and DC Hipot testing functions can be programmed over a range, 1µA to 40 mA AC and 0.1µA to 20 mA DC, providing flexibility in testing a range of devices. The analyzer's insulation resistance test (similar to DC Hipot test) calculates and displays a product's insulation resistance value in ohms at any programmed test voltage between 50 and 1000 V DC. The ground bond test is capable of supplying 1 to 30 A for verifying the integrity of a product's ground system. The earth leakage current test can be set to measure the current flowing through a product's line cord ground connector in three possible configurations; normal operating conditions, reverse line, or a single fault mode with one of a possible five human body models selected, all in accordance with UL standards. Additionally, the unit is capable of monitoring line voltage, current or power consumption (VA) during this earth leakage current test.